6.5. Registers for Write and Readout
Table 11
Values placed in 16-bit
registers
Write (w)/Readout (r)
Input type4000
tYP1
w/r 0...7
Range
Description
Symbol
0 Pulse counter
1 Frequency (f < 10 kHz)
2 Frequency (f > 10 kHz)
3 Rotational speed
4 Period
5 Long period
6 Worktime counter
7 Current time
8 Incremental encoder
Value
4003
Cnt1
w/r 1...3600
Averaged measurement time expressed in seconds.
This time defines the averaging time of the measured
value. The displayed value is the mean value
calculated from the Cnt1 period.
Selection of input quantity recalibration
4001
SCAL1
w/r 0, 1
Value
0 Multiplication by the constant
1 Division by the constant
Description
Permit for external functions Start/Stop.
Erasing
4002
E_In1
w/r 0...2
Value
0 External functions switched off. Access
to functions from the keyboard level
1 External functions switched on. Access
by means of buttons switched off
2 External functions switched on.
Access by means of the keyboard and
control inputs.